Drop coating deposition Raman (DCDR) spectroscopy of contaminants

Spectrochim Acta A Mol Biomol Spectrosc. 2021 Dec 5:262:120109. doi: 10.1016/j.saa.2021.120109. Epub 2021 Jun 24.

Abstract

Raman spectroscopy is a useful technique to identify small organic molecules, including contaminants. The drop coating deposition Raman (DCDR) is more sensitive than conventional Raman spectroscopy from solution. It is based on Raman measurement from a small drop dried on a hydrophobic surface where studied molecules are preconcentrated. In this paper, DCDR spectra of dried drops of selected contaminants (food contaminant melamine, fungicide thiram, herbicides bentazon and picloram) on the hydrophobic substrate were acquired for the first time, whereas Raman spectra from stock solutions were impossible to obtain under the same experimental conditions. The lowest DCDR detected concentrations were determined as 6.4 µM, 0.31 µM, 20 µM and 2 µM in deposited concentrations for melamine, thiram, bentazon and picloram, respectively. Therefore, DCDR spectroscopy can serve to detect these molecules in concentrations relevant in food/groundwater contaminations.

Keywords: Bentazon; Contaminant; DCDR; Drop coating deposition Raman; Melamine; Picloram; Thiram.

MeSH terms

  • Hydrophobic and Hydrophilic Interactions
  • Spectrum Analysis, Raman*