The gamma and X-ray emission intensities of 169Er were determined using radionuclidically pure 169Er. The activity of the 169Er source was standardized by the triple-to-double-coincidence ratio technique. Three independent measurements were performed to measure the emission intensities using calibrated high-purity germanium spectrometers. The efficiencies were computed with the Monte Carlo method and validated using several experimental measurements. Final results present a large uncertainty reduction compared to previous evaluations. The emission intensities per decay of 169Er are reported as 1.401(40).10-5 for the 109.8 keV line and 1.513(19).10-6 for the 118.2 keV line. The values obtained for the X-ray lines show large discrepancies with the reference values.
Keywords: Erbium-169; Gamma-ray emission intensity; Gamma-ray spectrometry; Standardization; X-ray emission intensity.