Immunocomplex capture fluorescence analysis (ICFA) which basic principle is same as Luminex crossmatch (LXM), could detect donor-specific HLA antibody (DSA). The advantages of ICFA are (i) detection of DSA and (ii) no requirement of viable cells over the flow cytometry crossmatch (FCXM). However, FCXM has been widely used because of its higher sensitivity than ICFA, in particular HLA-class II antibody detection. In this study the accuracy of DSA detection against HLA-class II was investigated by modifying the original method of ICFA. Increment of the sensitivity was found when purified peripheral blood mononuclear cells (PBMCs) were used instead of whole blood. An ICFA-PBMC in addition to FCXM-T/B was conducted for 118 patients before kidney transplantation and 13 patients with de novo DSA against HLA-class II after transplantation. Significantly positive correlation was observed between the values of ICFA-PBMC and DSA mean fluorescence intensity (MFI) targeting class II (p < 0.0001). When the cutoff level of 1.4 was determined by receiver operating characteristic curve analysis, the average DSA MFI was found to be significantly higher in the ICFA-PBMC (class II) positive group comparing to that in the negative group (12,217 vs 3885, p = 0.0027). ICFA-PBMC and optimized cutoff level could provide valid information in cases of suspected DSA.
Keywords: DSA; Flow cytometry cross match tests; Immunocomplex capture fluorescence analysis; Luminex crossmatch; Renal transplantation.