More than the Eye Can See: Shedding New Light on SARS-CoV-2 Lateral Flow Device-Based Immunoassays

ACS Appl Mater Interfaces. 2021 Jun 9;13(22):25694-25700. doi: 10.1021/acsami.1c04283. Epub 2021 May 28.

Abstract

Containing the global severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) pandemic has been an unprecedented challenge due to high horizontal transmissivity and asymptomatic carriage rates. Lateral flow device (LFD) immunoassays were introduced in late 2020 to detect SARS-CoV-2 infection in asymptomatic or presymptomatic individuals rapidly. While LFD technologies have been used for over 60 years, their widespread use as a public health tool during a pandemic is unprecedented. By the end of 2020, data from studies into the efficacy of the LFDs emerged and showed these point-of-care devices to have very high specificity (ability to identify true negatives) but inadequate sensitivity with high false-negative rates. The low sensitivity (<50%) shown in several studies is a critical public health concern, as asymptomatic or presymptomatic carriers may wrongly be assumed to be noninfectious, posing a significant risk of further spread in the community. Here, we show that the direct visual readout of SARS-CoV-2 LFDs is an inadequate approach to discriminate a potentially infective viral concentration in a biosample. We quantified significant immobilized antigen-antibody-labeled conjugate complexes within the LFDs visually scored as negative using high-sensitivity synchrotron X-ray fluorescence imaging. Correlating quantitative X-ray fluorescence measurements and quantitative reverse transcription-polymerase chain reaction (qRT-PCR) determined numbers of viral copies, we identified that negatively scored samples could contain up to 100 PFU (equivalent here to ∼10 000 RNA copies/test). The study demonstrates where the shortcomings arise in many of the current direct-readout SARS-CoV-2 LFDs, namely, being a deficiency in the readout as opposed to the potential level of detection of the test, which is orders of magnitude higher. The present findings are of importance both to public health monitoring during the Coronavirus Disease 2019 (COVID-19) pandemic and to the rapid refinement of these tools for immediate and future applications.

Keywords: SARS-CoV-2; X-ray fluorescence; antigen test; gold nanoparticles; immunogold; lateral flow device; sensitivity; specificity.

MeSH terms

  • Animals
  • COVID-19 / diagnosis*
  • COVID-19 / virology*
  • Chlorocebus aethiops
  • Immunoassay / instrumentation*
  • Microscopy, Electron, Transmission
  • Real-Time Polymerase Chain Reaction
  • Reference Standards
  • Sensitivity and Specificity
  • Severe acute respiratory syndrome-related coronavirus / isolation & purification*
  • Severe acute respiratory syndrome-related coronavirus / ultrastructure
  • Spectrometry, X-Ray Emission
  • Vero Cells