Alleviating carbohydrate counting with a FiASP-plus-pramlintide closed-loop delivery system (artificial pancreas): Feasibility and pilot studies

Diabetes Obes Metab. 2021 Sep;23(9):2090-2098. doi: 10.1111/dom.14447. Epub 2021 Jun 21.

Abstract

Aim: To assess whether a FiASP-and-pramlintide closed-loop system has the potential to replace carbohydrate counting with a simple meal announcement (SMA) strategy (meal priming bolus without carbohydrate counting) without degrading glycaemic control compared with a FiASP closed-loop system.

Materials and methods: We conducted a 24-hour feasibility study comparing a FiASP system with full carbohydrate counting (FCC) with a FiASP-and-pramlintide system with SMA. We conducted a subsequent 12-day outpatient pilot study comparing a FiASP-and-placebo system with FCC, a FiASP-and-pramlintide system with SMA, and a FiASP-and-placebo system with SMA. Basal-bolus FiASP-and-pramlintide were delivered at a fixed ratio (1 U:10 μg). Glycaemic outcomes were measured, surveys evaluated gastrointestinal symptoms and diabetes distress, and participant interviews helped establish a preliminary coding framework to assess user experience.

Results: Seven participants were included in the feasibility analysis. Time spent in 3.9-10 mmol/L was similar between both interventions (81%-84%). Four participants were included in the pilot analysis. Time spent in 3.9-10 mmol/L was similar between the FiASP-and-placebo with FCC and FiASP-and-pramlintide with SMA interventions (70%), but was lower in the FiASP-and-placebo with SMA intervention (60%). Time less than 3.9 mmol/L and gastrointestinal symptoms were similar across all interventions. Emotional distress was moderate at baseline, after the FiASP-and-placebo with FCC and SMA interventions, and fell after the FiASP-and-pramlintide with SMA intervention. SMA reportedly afforded participants flexibility and reduced mealtime concerns.

Conclusions: The FiASP-and-pramlintide system has the potential to substitute carbohydrate counting with SMA without degrading glucose control.

Keywords: artificial pancreas; carbohydrate counting; pramlintide; simple meal announcement; type 1 diabetes.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Blood Glucose
  • Diabetes Mellitus, Type 1* / drug therapy
  • Feasibility Studies
  • Humans
  • Hypoglycemic Agents / therapeutic use
  • Insulin / therapeutic use
  • Insulin Infusion Systems
  • Islet Amyloid Polypeptide / therapeutic use
  • Pancreas, Artificial*
  • Pilot Projects

Substances

  • Blood Glucose
  • Hypoglycemic Agents
  • Insulin
  • Islet Amyloid Polypeptide
  • pramlintide