Localized Bioimpedance Measurements with the MAX3000x Integrated Circuit: Characterization and Demonstration

Sensors (Basel). 2021 Apr 25;21(9):3013. doi: 10.3390/s21093013.

Abstract

The commercial availability of integrated circuits with bioimpedance sensing functionality is advancing the opportunity for practical wearable systems that monitor the electrical impedance properties of tissues to identify physiological features in support of health-focused applications. This technical note characterizes the performance of the MAX3000x (resistance/reactance accuracy, power modes, filtering, gains) and is available for on-board processing (electrode detection) for localized bioimpedance measurements. Measurements of discrete impedances that are representative of localized tissue bioimpedance support that this IC has a relative error of <10% for the resistance component of complex impedance measurements, but can also measure relative alterations in the 250 mΩ range. The application of the MAX3000x for monitoring localized bicep tissues during activity is presented to highlight its functionality, as well as its limitations, for multi-frequency measurements. This device is a very-small-form-factor single-chip solution for measuring multi-frequency bioimpedance with significant on-board processing with potential for wearable applications.

Keywords: Maxim MAX3000x; bioimpedance; localized tissues; multi-frequency; resistance/reactance.

MeSH terms

  • Electric Impedance*
  • Electrodes
  • Monitoring, Physiologic