Deep-learning-based direct synthesis of low-energy virtual monoenergetic images with multi-energy CT

J Med Imaging (Bellingham). 2021 Sep;8(5):052104. doi: 10.1117/1.JMI.8.5.052104. Epub 2021 Apr 19.

Abstract

Purpose: We developed a deep learning method to reduce noise and beam-hardening artifact in virtual monoenergetic image (VMI) at low x-ray energy levels. Approach: An encoder-decoder type convolutional neural network was implemented with customized inception modules and in-house-designed training loss (denoted as Incept-net), to directly estimate VMI from multi-energy CT images. Images of an abdomen-sized water phantom with varying insert materials were acquired from a research photon-counting-detector CT. The Incept-net was trained with image patches ( 64 × 64 pixels ) extracted from the phantom data, as well as synthesized, random-shaped numerical insert materials. The whole CT images ( 512 × 512 pixels ) with the remaining real insert materials that were unseen in network training were used for testing. Seven contrast-enhanced abdominal CT exams were used for preliminary evaluation of Incept-net generalizability over anatomical background. Mean absolute percentage error (MAPE) was used to evaluate CT number accuracy. Results: Compared to commercial VMI software, Incept-net largely suppressed beam-hardening artifact and reduced noise (53%) in phantom study. Incept-net presented comparable CT number accuracy at higher-density ( P -value [0.0625, 0.999]) and improved it at lower-density inserts ( P - value = 0.0313 ) with overall MAPE: Incept-net [2.9%, 4.6%]; commercial-VMI [6.7%, 10.9%]. In patient images, Incept-net suppressed beam-hardening artifact and reduced noise (up to 50%, P - value = 0.0156 ). Conclusion: In this preliminary study, Incept-net presented the potential to improve low-energy VMI quality.

Keywords: artifact reduction; convolutional neural network; deep learning; dual-energy CT; noise reduction; photon counting detector; virtual monoenergetic image.