High-power, short-duration ablation in the coronary sinus: clinical cases and preliminary observations on swine hearts

J Interv Card Electrophysiol. 2022 Mar;63(2):311-321. doi: 10.1007/s10840-021-00994-0. Epub 2021 Apr 15.

Abstract

Purpose: Coronary sinus-related arrhythmias are common; however, it is difficult to perform radiofrequency (RF) ablation at these sites efficiently and safely. High-power, short-duration ablation (HPSD) is a proven alternative strategy for pulmonary vein isolation (PVI); whether it can be applied to ablation of the coronary sinus is unknown. The purpose of this preliminary study was to evaluate the feasibility and safety of HPSD ablation in the coronary sinus.

Methods: Firstly, we demonstrated 4 clinical cases of 3 types of arrhythmias who had unsuccessful ablation with standard power initially, but received successful ablations with HPSD. Secondly, RF ablation was performed in the coronary sinus ostium (CSO) and middle cardiac vein (MCV) of 4 in vitro swine hearts. Two protocols were compared: HPSD (45 W/5 S×5 rounds) and a conventional strategy that used low-power, long-duration ablation (LPLD: 25 W/10 S ×5 rounds). The total duration of HPSD protocol was 25 s, and which of LPLD was 50 s.

Results: A total of 28 lesions were created. HPSD can produce longer, wider, deeper, and larger lesions than LPLD. This difference was more pronounced when the ablation was in the MCV. One instance of steam pop occurred during LPLD in the MCV.

Conclusions: HPSD is an effective alternative strategy for ablation in coronary sinus according to clinical applications and preliminary animal study. However, the safety needs to be further evaluated based on more animal and clinical studies.

Keywords: Atrial fibrillation; Coronary sinus; High-power and short-duration; Radiofrequency ablation.

MeSH terms

  • Animals
  • Atrial Fibrillation* / surgery
  • Catheter Ablation* / methods
  • Coronary Sinus* / surgery
  • Humans
  • Pulmonary Veins* / surgery
  • Swine
  • Treatment Outcome