Background: Highly localized impedance (LI) measurements during atrial fibrillation (AF) ablation have recently emerged as a viable real-time indicator of tissue characteristics and durability of the lesions created. We report the outcomes of acute and long-term clinical evaluation of the new DirectSense algorithm in AF ablation.
Methods: Consecutive patients undergoing AF ablation were included in the CHARISMA registry. RF delivery was guided by the DirectSense algorithm, which records the magnitude and time-course of the impedance drop. The ablation endpoint was pulmonary vein isolation (PVI), as assessed by the entrance and exit block.
Results: 3556 point-by-point first-pass RF applications of >10 s duration were analyzed in 153 patients (mean age=59 ± 10 years, 70% men, 61% paroxysmal AF, 39% persistent AF). The mean baseline LI was 105 ± 15 Ω before ablation and 92 ± 12 Ω after ablation (p < .0001). Both absolute drops in LI and the time to LI drop (LI drop/τ) were greater at successful ablation sites (n = 3122, 88%) than at ineffective ablation sites (n = 434, 12%) (14 ± 8 Ω vs 6 ± 4 Ω, p < .0001 for LI; 0.73 [0.41-1.25] Ω/s vs. 0.35[0.22-0.59 Ω/s, p < .0001 for LI drop/τ). No major complications occurred during or after the procedures. All PVs had been successfully isolated. During a mean follow-up of 366 ± 130 days, 18 patients (11.8%) suffered an AF/atrial tachycardia recurrence after the 90-day blanking period.
Conclusion: The magnitude and time-course of the LI drop during RF delivery were associated with effective lesion formation. This ablation strategy for PVI guided by LI technology proved safe and effective and resulted in a very low rate of AF recurrence over 1-year follow-up.
Keywords: DirectSense; atrial fibrillation; catheter ablation; lesion formation; local impedance.