Background: For peripheral artery disease (PAD), MR angiography (MRA) is a well-established diagnostic modality providing morphologic and dynamic information comparable to digital subtraction angiography (DSA). However, relatively large amounts of contrast agents are necessary to achieve this.
Purpose: To evaluate the diagnostic accuracy of time-resolved 4D MR-angiography with interleaved stochastic trajectories (TWIST-MRA) by using maximum intensity projections (MIPs) of dynamic images acquired with reduced doses of contrast agent.
Study type: Retrospective.
Population: Forty adult PAD patients yielding 1088 artery segments.
Field strength/sequence: A 3.0 T, time-resolved 4D MR-angiography with TWIST-MRA and MIP of dynamic images.
Assessment: DSA was available in 14 patients (256 artery segments) and used as reference standard. Three-segmental MIP reconstructions of TWIST-images after administration of 3 mL of gadolinium-based contrast agent (Gadoteridol/Prohance®, 0.5 M) per anatomical level (pelvis, thighs, and lower legs) yielded 256 artery segments for correlation between MRA and DSA. Three independent observers rated image quality (scale: 1 [nondiagnostic] to 4 [excellent]) and the degree of venous overlay (scale: 0 [none] to 2 [significant]) for all segments. Diagnostic accuracy for the detection of >50% stenosis and artery occlusion was calculated for all observers.
Statistical tests: Binary classification test (sensitivity, specificity, positive/negative predictive values, diagnostic accuracy). Intraclass correlation coefficients (ICCs), logistic regression analysis with comparison of areas under the receiver-operating-characteristics (ROC) curves (AUCs) with the DeLong method. Bland-Altman-comparison.
Results: High diagnostic performance was achieved for the detection of >50% stenosis (sensitivity 92.9% [84.3-99.9% (95%-CI)] and specificity 98.5% [95.7-99.8% (95%-CI)]) and artery occlusion (sensitivity 93.1% [77.2-99.2% (95%-CI)] and specificity 99.1% [96.9-99.9% (95%-CI)]). Inter-reader agreement was excellent with ICC values ranging from 0.95 to 1.0 for >50% artery stenosis and occlusion. Image quality was good to excellent for both readers (3.41 ± 0.72, 3.33 ± 0.65, and 3.38 ± 0.61 [mean ± SD]) with good correlation between observer ratings (ICC 0.71-0.81). No significant venous overlay was observed (0.06 ± 0.24, 0.23 ± 0.43 and 0.11 ± 0.45 [mean ± SD]).
Data conclusion: MIPs of dynamic TWIST-MRA offer a promising diagnostic alternative necessitating only reduced amounts (50%) of gadolinium-based contrast agents for the entire runoff vasculature.
Evidence level: 3 TECHNICAL EFFICACY: Stage 2.
Keywords: digital subtraction angiography; peripheral arterial disease; time-resolved MR angiography; time-resolved angiography with interleaved stochastic trajectories.
© 2021 The Authors. Journal of Magnetic Resonance Imaging published by Wiley Periodicals LLC. on behalf of International Society for Magnetic Resonance in Medicine.