Amperometric aptasensor with sandwich-type architecture for troponin I based on carboxyethylsilanetriol-modified graphene oxide coated electrodes

Biosens Bioelectron. 2021 Jul 1:183:113203. doi: 10.1016/j.bios.2021.113203. Epub 2021 Mar 29.

Abstract

A novel amperometric aptasensor for the specific detection of cardiac troponin I (cTnI) was constructed by using screen-printed carbon electrodes coated with a carboxyethylsilanetriol-modified graphene oxide derivative as transduction element. This novel carboxylic acid-enriched nanomaterial allows easy and high load immobilization of the capture aptamer molecules on the electrode surface. The biosensing interface was assembled by covalent attachment of an amino-functionalized DNA aptamer on the carboxylic acid-enriched electrode surface. The sensing approach relies on the specific recognition of cTnI by the aptamer and further assembly of a sandwich-type architecture with a novel aptamer-peroxidase conjugate as signaling element. The aptasensor was employed to detect the cardiac biomarker in the broad range from 1.0 pg/mL to 1.0 μg/mL with a detection limit of 0.6 pg/mL. This electroanalytical device also showed high specificity, reproducibility and stability, and was useful to quantify cTnI in reconstituted human serum samples.

Keywords: Aptamer; Biosensor; Cardiac troponin I; Graphene; Screen-printed electrode.

MeSH terms

  • Aptamers, Nucleotide*
  • Biosensing Techniques*
  • Electrochemical Techniques
  • Electrodes
  • Gold
  • Graphite*
  • Humans
  • Limit of Detection
  • Reproducibility of Results
  • Troponin I

Substances

  • Aptamers, Nucleotide
  • Troponin I
  • graphene oxide
  • Gold
  • Graphite