Abstract

Passive micromixers are miniaturized instruments that are used to mix fluids in microfluidic systems. In microchannels, combination of laminar flows and small diffusion constants of mixing liquids produce a difficult mixing environment. In particular, in very low Reynolds number flows, e.g., Re < 10, diffusive mixing cannot be promoted unless a large interfacial area is formed between the fluids to be mixed. Therefore, the mixing distance increases substantially due to a slow diffusion process that governs fluid mixing. In this article, a novel 3-D passive micromixer design is developed to improve fluid mixing over a short distance. Computational Fluid Dynamics (CFD) simulations are used to investigate the performance of the micromixer numerically. The circular-shaped fluid overlapping (CSFO) micromixer design proposed is examined in several fluid flow, diffusivity, and injection conditions. The outcomes show that the CSFO geometry develops a large interfacial area between the fluid bodies. Thus, fluid mixing is accelerated in vertical and/or horizontal directions depending on the injection type applied. For the smallest molecular diffusion constant tested, the CSFO micromixer design provides more than 90% mixing efficiency in a distance between 260 and 470 µm. The maximum pressure drop in the micromixer is found to be less than 1.4 kPa in the highest flow conditioned examined.

Keywords: CFD; concentric flow; diffusive mixing; fluid overlapping; micromixer; passive mixing; segmentation; sequential injection.