Accuracy of Pedicle Screw Placement Comparing an Electronic Conductivity Device and a Multi-axis Angiography Unit with C-arm Fluoroscopy in Lumbar Fixation Surgery for Safety

Neurol Med Chir (Tokyo). 2021 May 15;61(5):334-340. doi: 10.2176/nmc.oa.2020-0374. Epub 2021 Mar 31.

Abstract

The aim of this study was to compare the accuracy, safety, and usefulness of percutaneous pedicle screw (PPS) placement for lumbar fixation using a multi-axis angiography unit (MAU) and an electronic conductivity device (ECD) with a cannulated Jamshidi needle with that using a conventional C-arm. Of 65 cases that underwent lumbar fixation (region between L1-S1) during April 2013 to March 2019, 57 cases that could be followed-up for more than 12 months after the procedure were included. Among them, 31 patients (150 screws) received treatment with MAU and ECD (MAU+ECD group) and 26 (117 screws) were treated with the conventional C-arm. We performed a retrospective study of the surgical techniques used in each group at our institute by assessing the accuracy of PPS using Gertzbin-Robbins classification and the Japanese Orthopedic Association (JOA) score for recovery. There was no significant difference in surgery outcome based on the JOA recovery rate. There was a significant difference between the two groups in terms of Accuracy-1 (Group A indicating accuracy and Groups B-E indicating inaccuracy), where the rates were 85.3% and 72.0% in the MAU+ECD group and C-arm group, respectively (P = 0.008). There was also a significant difference between the two groups in terms of Accuracy-2 (Groups A-B indicating accuracy; Groups C-E indicate inaccuracy), where the rates were 98.0% and 92.4% in the MAU+ECD and C-arm groups, respectively (P = 0.036). A combination of MAU and ECD is a safe and accurate method for inserting screws into the pedicle.

Keywords: accuracy rate; electronic conductivity device; lumbar fixation surgery; multi-axis angiography unit; percutaneous pedicle screw.

MeSH terms

  • Angiography
  • Electronics
  • Fluoroscopy
  • Humans
  • Lumbar Vertebrae / diagnostic imaging
  • Lumbar Vertebrae / surgery
  • Pedicle Screws*
  • Retrospective Studies
  • Spinal Fusion*