Use of Phil Embolic Agent for Bleeding in Non-Neurological Interventions

J Clin Med. 2021 Feb 11;10(4):701. doi: 10.3390/jcm10040701.

Abstract

Objective: To evaluate the safety and efficacy of the Phil liquid embolic agent in non-neurological embolization procedures. M&M: Thirty-five patients with a mean age of 62.5 years underwent percutaneous embolization using Phil for the treatment of visceral arterial bleedings in 20/35 patients (including three gluteal, one bladder, two superior mesenteric, three epigastric, one deep femoral, five internal iliac, four intercostal, and one lingual arteries), splanchnic pseudoaneurysms in 11/35 patients (including three hepatic, five splenic, and three renal arteries), pancreatic bleeding metastasis in 1/35 patient, and gastric bleeding varices in 3/35 patients. Phil is composed of a non-adhesive copolymer dissolved in DMSO (Anhydrous Dimethyl Sulfoxide) with different viscosity. Procedures were performed slowly under continuous fluoroscopic guidance to avoid embolization of non-target vessels.

Results: Clinical success was obtained with a single intervention in 34 cases (97.15%), while a repeated procedure was required in one case (2.85%). No technical complications nor non-target embolization occurred. A case of post-embolic syndrome was noted (2.85%) in one patient. DMSO administration-related pain was successfully controlled by medical therapy.

Conclusion: Phil can be considered a safe and effective embolic agent for the treatment of non-neurologic bleeding.

Keywords: bleeding; liquid embolic agent; non-neurologic intervention; transcatheter embolization.