A robotically assisted 3D printed quality assurance lung phantom for Calypso

Phys Med Biol. 2021 Apr 6;66(7). doi: 10.1088/1361-6560/abebaa.

Abstract

Purpose. Radiation dose delivered to targets located near the upper-abdomen or in the thorax are significantly affected by respiratory-motion. Relatively large-margins are commonly added to compensate for this motion, limiting radiation-dose-escalation. Internal-surrogates of target motion, such as a radiofrequency (RF) tracking system, i.e. Calypso®System, are used to overcome this challenge and improve normal-tissue sparing. RF tracking systems consist of implanting transponders in the vicinity of the tumor to be tracked using radiofrequency-waves. Unfortunately, although the manufacture provides a universal quality-assurance (QA) phantom, QA-phantoms specifically for lung-applications are limited, warranting the development of alternative solutions to fulfil the tests mandated by AAPM's TG142. Accordingly, our objective was to design and develop a motion-phantom to evaluate Calypso for lung-applications that allows the Calypso®Beacons to move in different directions to better simulate truelung-motion.Methods and Materials.A Calypso lung QA-phantom was designed, and 3D-printed. The design consists of three independent arms where the transponders were attached. A pinpoint-chamber with a buildup-cap was also incorporated. A 4-axis robotic arm was programmed to drive the motion-phantom to mimic breathing. After acquiring a four-dimensional-computed-tomography (4DCT) scan of the motion-phantom, treatment-plans were generated and delivered on a Varian TrueBeam®with Calypso capabilities. Stationary and gated-treatment plans were generated and delivered to determine the dosimetric difference between gated and non-gated treatments. Portal cine-images were acquired to determine the temporal-accuracy of delivery by calculating the difference between the observed versus expected transponders locations with the known speed of the transponders' motion.Results.Dosimetric accuracy is better than the TG142 tolerance of 2%. Temporal accuracy is greater than, TG142 tolerance of 100 ms for beam-on, but less than 100 ms for beam-hold.Conclusions.The robotic QA-phantom designed and developed in this study provides an independent phantom for performing Calypso lung-QA for commissioning and acceptance testing of Calypso for lung treatments.

Keywords: 3D printing; Calypso; motion management; quality assurance; robotics.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Humans
  • Lung / diagnostic imaging
  • Neonicotinoids
  • Phantoms, Imaging
  • Printing, Three-Dimensional
  • Robotic Surgical Procedures*
  • Thiazines

Substances

  • Neonicotinoids
  • Thiazines
  • thiacloprid