Migration behaviour of 2 clinically excellent cementless stems with different design rationales: 5-year follow-up of a randomised RSA-study

Hip Int. 2022 Nov;32(6):747-758. doi: 10.1177/1120700021995482. Epub 2021 Feb 17.

Abstract

Introduction: Excellent long-term survival has been reported for both the Taperloc and the Mallory-Head cementless stems. However, little is known about the migration behaviour of these stems which have different design rationales. The purpose of this randomised clinical trial was to compare the migration and clinical outcomes of these stems during 5 years of follow-up.

Methods: 42 consecutive hips in 38 patients scheduled to receive cementless THA were randomised to either a Taperloc or a Mallory-Head stem. Evaluation took place preoperatively and postoperatively on the second day, at 6, 12, 26, and 52 weeks, and annually thereafter. Primary outcome was stem migration measured using roentgen stereophotogrammetric analysis (RSA) and secondary outcomes were the Harris Hip Score (HHS) and 36-Item Short-Form Health Survey (SF-36). No patients were lost to follow-up; in 1 patient the THA was removed due to deep infection 3 months postoperatively. In 6 hips migration measurements were not possible due to insufficient marker configuration.

Results: Throughout the follow-up period of 5 years, 3-dimensional migration was comparable between the Taperloc and the Mallory-Head stems (p-values > 0.05). However, at the 5-year follow-up point the retroversion of the Mallory-Head stem was 0.9° more than the Taperloc stem (p = 0.04). Initial subsidence and retroversion were respectively as large as 6.8 mm and 3.6° for the Taperloc stem and 5 mm and 3.6° for the Mallory-Head stem. After the first postoperative year, both implants had stabilised. The mean increment of HHS, as well as the SF-36 scores during the 5-year follow-up, were comparable between the 2 stems.

Conclusions: The excellent long-term survival of both designs was confirmed in this study showing comparable initial migration with subsequent stabilisation. However, the Taperloc design with a flat, wedged geometry showed better rotational stability.

Keywords: Randomised clinical trial; Roentgen stereophotogrammetric analysis; total hip arthroplasty.

Publication types

  • Randomized Controlled Trial

MeSH terms

  • Arthroplasty, Replacement, Hip* / methods
  • Follow-Up Studies
  • Hip Prosthesis*
  • Humans
  • Prosthesis Design
  • Prosthesis Failure
  • Radiostereometric Analysis
  • Treatment Outcome