Comparing the accuracy of transcutaneous sensor and 90-day implantable glucose sensor

Nutr Metab Cardiovasc Dis. 2021 Feb 8;31(2):650-657. doi: 10.1016/j.numecd.2020.09.006. Epub 2020 Sep 12.

Abstract

Background and aims: Continuous glucose monitoring improves glycemic control in diabetes. This study compared the accuracy of the Dexcom G5 Mobile (Dexcom, San Diego, CA) transcutaneous sensor (DG5) and the first version of Eversense (Senseonics,Inc., Germantown, MD) implantable sensor (EVS).

Methods and results: Subjects with type 1 diabetes (T1D) and using EVS wore simultaneously DG5 for seven days. At day 3, patients were admitted to a clinical research center (CRC) to receive breakfast with delayed and increased insulin bolus to induce glucose excursions. At CRC, venous glucose was monitored every 15 min (or 5 min during hypoglycemia) for 6 h by YSI 2300 STAT PLUS™ glucose and lactate analyzer. At home patients were requested to perform 4 fingerstick glucose measurements per day. Eleven patients (9 males, age 47.4 ± 11.3 years, M±SD) were enrolled. During home-stay the median [25th-75th percentile] absolute relative difference (ARD) over all CGM-fingerstick matched-pairs was 11.64% [5.38-20.65]% for the DG5 and 10.75% [5.15-19.74]% for the EVS (p-value = 0.58). At CRC, considering all the CGM-YSI matched-pairs, the DG5 showed overall smaller median ARD than EVS, 7.91% [4.14-14.30]% vs 11.4% [5.04-18.54]% (p-value<0.001). Considering accuracy during blood glucose swings, DG5 performed better than EVS when glucose rate-of-change was -0.5 to -1.5 mg/dL/min, with median ARD of 7.34% [3.71-12.76]% vs 13.59% [4.53-20.78]% (p-value<0.001), and for rate-of-change < -1.5 mg/dl/min, with median ARD of 5.23% [2.09-15.29]% vs 12.73% [4.14-20.82]% (p-value = 0.02).

Conclusions: DG5 was more accurate than EVS at CRC, especially when glucose decreased. No differences were found at home.

Keywords: Accuracy; Continuous glucose monitoring; Hypoglycemia; Type 1 diabetes.

Publication types

  • Clinical Trial
  • Comparative Study

MeSH terms

  • Adult
  • Biomarkers / blood
  • Blood Glucose / drug effects
  • Blood Glucose / metabolism*
  • Blood Glucose Self-Monitoring / instrumentation*
  • Diabetes Mellitus, Type 1 / blood
  • Diabetes Mellitus, Type 1 / diagnosis*
  • Diabetes Mellitus, Type 1 / therapy
  • Equipment Design
  • Female
  • Glycemic Control
  • Humans
  • Hypoglycemic Agents / therapeutic use
  • Insulin / therapeutic use
  • Male
  • Middle Aged
  • Predictive Value of Tests
  • Reproducibility of Results
  • Time Factors
  • Transducers*
  • Treatment Outcome
  • Wireless Technology / instrumentation*

Substances

  • Biomarkers
  • Blood Glucose
  • Hypoglycemic Agents
  • Insulin