Automatic treatment planning for VMAT-based total body irradiation using Eclipse scripting

J Appl Clin Med Phys. 2021 Mar;22(3):119-130. doi: 10.1002/acm2.13189. Epub 2021 Feb 10.

Abstract

The purpose of this work is to establish an automated approach for a multiple isocenter volumetric arc therapy (VMAT)-based TBI treatment planning approach. Five anonymized full-body CT imaging sets were used. A script was developed to automate and standardize the treatment planning process using the Varian Eclipse v15.6 Scripting API. The script generates two treatment plans: a head-first VMAT-based plan for upper body coverage using four isocenters and a total of eight full arcs; and a feet-first AP/PA plan with three isocenters that covers the lower extremities of the patient. PTV was the entire body cropped 5 mm from the patient surface and extended 3 mm into the lungs and kidneys. Two plans were generated for each case: one to a total dose of 1200 cGy in 8 fractions and a second one to a total dose of 1320 cGy in 8 fractions. Plans were calculated using the AAA algorithm and 6 MV photon energy. One plan was created and delivered to an anthropomorphic phantom containing 12 OSLDs for in-vivo dose verification. For the plans prescribed to 1200 cGy total dose the following dosimetric results were achieved: median PTV V100% = 94.5%; median PTV D98% = 89.9%; median lungs Dmean = 763 cGy; median left kidney Dmean = 1058 cGy; and median right kidney Dmean = 1051 cGy. For the plans prescribed to 1320 cGy total dose the following dosimetric results were achieved: median PTV V100% = 95.0%; median PTV D98% = 88.7%; median lungs Dmean = 798 cGy; median left kidney Dmean = 1059 cGy; and median right kidney Dmean = 1064 cGy. Maximum dose objective was met for all cases. The dose deviation between the treatment planning dose and the dose measured by the OSLDs was within ±4%. In summary, we have demonstrated that scripting can produce high-quality plans based on predefined dose objectives and can decrease planning time by automatic target and optimization contours generation, plan creation, field and isocenter placement, and optimization objectives setup.

Keywords: advanced treatment planning; eclipse scripting; plan automation; special procedures; total body irradiation (TBI).

MeSH terms

  • Humans
  • Radiometry
  • Radiotherapy Dosage
  • Radiotherapy Planning, Computer-Assisted
  • Radiotherapy, Intensity-Modulated*
  • Whole-Body Irradiation*