Contrast-Enhanced Ultrasound of the Pleural Cavity: A Method to Locate Pleural Catheters and Identify Fibrous Septa

Ultrasound Med Biol. 2021 May;47(5):1261-1268. doi: 10.1016/j.ultrasmedbio.2021.01.011. Epub 2021 Feb 1.

Abstract

We aimed to explore the value of contrast-enhanced ultrasound (CEUS) of the pleural cavity in locating catheters and identifying fibrous septa and to compare CEUS with multiple existing methods. We included 304 participants whose pleural effusion could not continue to be drained and compared the catheter-localization capabilities of empirical diagnosis, B-mode ultrasound with normal saline and CEUS, with computed tomography as the reference standard. CEUS performed the best (accuracy, 100%; sensitivity, 100%; specificity, 100%), followed by B-mode ultrasound with normal saline (accuracy, 77.78%; sensitivity, 62.5%; specificity, 100%), and finally empirical diagnosis (accuracy, 54.17%; sensitivity, 66.67%; specificity, 33.33%). The capabilities of CEUS and computed tomography to identify fibrous septa were evaluated, with B-mode ultrasound as the reference, and CEUS (accuracy, 100%; sensitivity, 100%; specificity, 100%) was superior to computed tomography (accuracy, 82.41%; sensitivity, 26.09%; specificity, 97.65%). Overall, CEUS can accurately locate catheters and identify fibrous septa, with performance superior to existing methods.

Keywords: Catheter; Contrast-enhanced ultrasound; Pleural effusion; Septa; Ultrasound imaging.

Publication types

  • Comparative Study
  • Randomized Controlled Trial
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Aged
  • Catheters
  • Contrast Media*
  • Double-Blind Method
  • Female
  • Humans
  • Male
  • Middle Aged
  • Pleural Cavity / diagnostic imaging*
  • Prospective Studies
  • Ultrasonography / methods

Substances

  • Contrast Media