Comparative analysis of stress distribution in one-piece and two-piece implants with narrow and extra-narrow diameters: A finite element study

PLoS One. 2021 Feb 4;16(2):e0245800. doi: 10.1371/journal.pone.0245800. eCollection 2021.

Abstract

Objectives: The aim of this in vitro study was to evaluate the stress distribution on three implant models with narrow and extra-narrow diameters using the finite element method (FEA).

Materials and methods: Dental implants of extra-narrow diameter of 2.5 mm for a one-piece implant (group G1), a narrow diameter of 3.0 mm for a one-piece implant (group G2) and a narrow diameter of 3.5 mm for a two-piece implant with a Morse taper connection (group G3). A three-dimensional model was designed with cortical and cancellous bone, a crown and an implant/abutment set of each group. Axial and angled (30°) loads of 150 N was applied. The equivalent von Mises stress was used for the implants and peri-implant bone plus the Mohr-Coulomb analysis to confirm the data of the peri-implant bone.

Results: In the axial load, the maximum stress value of the cortical bone for the group G1 was 22.35% higher than that the group G2 and 321.23% than the group G3. Whereas in angled load, the groups G1 and G2 showing a similar value (# 3.5%) and a highest difference for the group G3 (391.8%). In the implant structure, the group G1 showed a value of 2188MPa, 93.6% higher than the limit.

Conclusions: The results of this study show that the extra-narrow one-piece implant should be used with great caution, especially in areas of non-axial loads, whereas the one- and two-piece narrow-diameter implants show adequate behavior in both directions of the applied load.

Publication types

  • Comparative Study

MeSH terms

  • Biomechanical Phenomena
  • Cancellous Bone
  • Cortical Bone
  • Crowns*
  • Finite Element Analysis*
  • Mechanical Tests*
  • Stress, Mechanical*
  • Weight-Bearing

Grants and funding

Biotecnos Research Center provided support in the form of salary for author SAG, who is the director and a researcher at the laboratory. The specific role of this author is articulated in the ‘author contributions’ section. Biotecnos Research Center is a company dedicated to conducting material analysis. The funder did not have any additional role in the study design, data collection and analysis, decision to publish, or preparation of the manuscript.