1.5-Dimensional Circular Array Transducer for In Vivo Endoscopic Ultrasonography

IEEE Trans Biomed Eng. 2021 Oct;68(10):2930-2939. doi: 10.1109/TBME.2021.3056140. Epub 2021 Sep 20.

Abstract

Objective: Traditional endoscopic ultrasonography (EUS), which uses one-dimensional (1-D) curvilinear or radial/circular transducers, cannot achieve dynamic elevational focusing, and the slice thickness is not sufficient. The purpose of this study was to design and fabricate a 1.5-dimensional (1.5-D) circular array transducer to achieve dynamic elevational focusing in EUS in vivo.

Methods: An 84 × 5 element 1.5-D circular array transducer was successfully developed and characterized in this study. It was fabricated with PZT-5H 1-3 composite that attained a high-electromechanical coupling factor and low-acoustic impedance. The acoustic field distribution was measured with different transmission modes to validate the 1.5-D elevational beam focusing capability. The imaging performance of the 84 × 5 element 1.5-D circular array transducer was evaluated by two wire phantoms, an agar-based cyst phantom, an ex vivo swine pancreas, and an in vivo rhesus macaque rectum based on multifocal ray-line imaging method with five-row elevational beam steering.

Results: It was demonstrated that the transducer exhibited a central frequency of 6.47 MHz with an average bandwidth of 50%, a two-way insertion loss of 23 dB, and crosstalk of <-26 dB around the center frequency.

Conclusion: Dynamic elevational focusing and the enhancement of the slice thickness in EUS were obtained with a 1.5-D circular array transducer.

Significance: This study promotes the development of multirow and two-dimensional array EUS probes for a more precise clinical diagnosis and treatment.