Course of Metal Ions after a Revision of Malfunctioning Metal-on-Metal Total Hip Prostheses

Medicina (Kaunas). 2021 Jan 28;57(2):115. doi: 10.3390/medicina57020115.

Abstract

The present research evaluated the course of cobalt and chromium in the blood and urine after the revision of metal-on-metal with a ceramic-on-polyethylene total hip arthroplasty. Seven patients were enrolled for hip prosthesis revision owing to ascertained damage of the implant. Metals in the blood and urine were evaluated before and after the hip revision. The double measurement before the total hip revision revealed high levels of metal ions (on average, 88.1 µg/L of cobalt in the blood, 399.0 µg/g of creatinine cobalt in the urine, 46.8 µg/L of chromium in the blood, and 129.6 µg/g of creatinine chromium in the urine at the first measurements), with an increasing trend between the first and second dosage. Within a week after the hip revision, the levels of metal ions significantly decreased by approximately half. Four to six months after the operation, the cobalt levels were found near to the reference values, whereas the chromium levels reached 25% of the values measured before the revision. The revision of malfunctioning metal-on-metal implants produced a dramatic decrease of metal ions in biological fluids, although it did not completely rescue the chromium level.

Keywords: chromium; cobalt; hip revision; metal ions; metal-on-metal hip prostheses.