Comparison of procedural efficacy and biophysical parameters between two competing cryoballoon technologies for pulmonary vein isolation: Insights from an initial multicenter experience

J Cardiovasc Electrophysiol. 2021 Mar;32(3):580-587. doi: 10.1111/jce.14915. Epub 2021 Feb 1.

Abstract

Introduction: Recently a novel cryoballoon system (POLARx, Boston Scientific) became available for the treatment of atrial fibrillation. This cryoballoon is comparable with Arctic Front Advance Pro (AFA-Pro, Medtronic), however, it maintains a constant balloon pressure. We compared the procedural efficacy and biophysical characteristics of both systems.

Methods: One hundred and ten consecutive patients who underwent first-time cryoballoon ablation (POLARx: n = 57; AFA-Pro: n = 53) were included in this prospective cohort study.

Results: Acute isolation was achieved in 99.8% of all pulmonary veins (POLARx: 99.5% vs. AFA-Pro: 100%, p = 1.00). Total procedure time (81 vs. 67 min, p < .001) and balloon in body time (51 vs. 35 min, p < .001) were longer with POLARx. After a learning curve, these times were similar. Cryoablation with POLARx was associated with shorter time to balloon temperature -30°C (27 vs. 31 s, p < .001) and -40°C (32 vs. 54 s, p < .001), lower balloon nadir temperature (-55°C vs. -47°C, p < .001), and longer thawing time till 0°C (16 vs. 9 s, p < .001). There were no differences in time-to-isolation (TTI; POLARx: 45 s vs. AFA-Pro 43 s, p = .441), however, POLARx was associated with a lower balloon temperature at TTI (-46°C vs. -37°C, p < .001). Factors associated with acute isolation differed between groups. The incidence of phrenic nerve palsy was comparable (POLARx: 3.5% vs. AFA-Pro: 3.7%).

Conclusion: The novel cryoballoon is comparable to AFA-Pro and requires only a short learning curve to get used to the slightly different handling. It was associated with faster cooling rates and lower balloon temperatures but TTI was similar to AFA-Pro.

Keywords: atrial fibrillation; catheter ablation; cryoablation; cryoballoon; pulmonary vein isolation.

Publication types

  • Multicenter Study

MeSH terms

  • Atrial Fibrillation* / diagnosis
  • Atrial Fibrillation* / surgery
  • Boston
  • Catheter Ablation*
  • Cryosurgery* / adverse effects
  • Humans
  • Prospective Studies
  • Pulmonary Veins* / diagnostic imaging
  • Pulmonary Veins* / surgery
  • Treatment Outcome