The 18F phantom clinical trials qualification for 18F-FDG-PET scanning adopted by GELTAMO (Grupo Español de Linfomas/Trasplante Autólogo de Médula Ósea)

Rev Esp Med Nucl Imagen Mol (Engl Ed). 2021 May-Jun;40(3):149-154. doi: 10.1016/j.remn.2020.06.003. Epub 2021 Jan 21.
[Article in English, Spanish]

Abstract

Introduction and objectives: Since different PET/CT (Positron Emission Tomography/Computed Tomography) scanners give different qualitative readings, a program for clinical trial qualification (CTQ) is mandatory to guarantee a reliable and reproducible use of PET/CT in prospective multi-centre clinical trials. Within this work we will show the results carried out in performing CTQ in Spain.

Materials and methods: We set up, under the auspices of Grupo Español de Linfomas/Trasplante Autólogo de Médula Osea (GELTAMO), a CTQ program consisting of the acquisition and analysis of 18F uniformity and image quality phantoms for the reduction of inter-scanner variability (ISV). The ISV was estimated on background activity concentration (BAC) and sphere to background ratio (SBR) and defined as their 95% confidence level.

Results: Twenty-six out of 27 (96%) scanners fulfilled the CTQ requirements. The CTQ was fulfilled at the first round in 27% of the cases, while in 38%, 15% and 20%, two, three or more than three iterations, were required, respectively. The mean CTQ time was (1.8 ± 1.4) months (range: 0.3-4.6). The ISV in BAC and SBR were 20.3% and 67.7%.

Conclusions: The CTQ proven to be a reliable tool to reduce ISV. This enabled to set-up clinical trials in which PET/CT was used to evaluate different clinical endpoints.

Keywords: (18)F-FDG; Armonización de escáneres; Linfoma; Lymphoma; PET/CT; PET/CT reproducibility; PET/TC; Reproducibilidad de la PET/TC; Scanner harmonization.