Novel methodologies for dosimetry audits: Adapting to advanced radiotherapy techniques

Phys Imaging Radiat Oncol. 2018 Mar 19:5:76-84. doi: 10.1016/j.phro.2018.03.002. eCollection 2018 Jan.

Abstract

With new radiotherapy techniques, treatment delivery is becoming more complex and accordingly, these treatment techniques require dosimetry audits to test advanced aspects of the delivery to ensure best practice and safe patient treatment. This review of novel methodologies for dosimetry audits for advanced radiotherapy techniques includes recent developments and future techniques to be applied in dosimetry audits. Phantom-based methods (i.e. phantom-detector combinations) including independent audit equipment and local measurement equipment as well as phantom-less methods (i.e. portal dosimetry, transmission detectors and log files) are presented and discussed. Methodologies for both conventional linear accelerator (linacs) and new types of delivery units, i.e. Tomotherapy, stereotactic devices and MR-linacs, are reviewed. Novel dosimetry audit techniques such as portal dosimetry or log file evaluation have the potential to allow parallel auditing (i.e. performing an audit at multiple institutions at the same time), automation of data analysis and evaluation of multiple steps of the radiotherapy treatment chain. These methods could also significantly reduce the time needed for audit and increase the information gained. However, to maximise the potential, further development and harmonisation of dosimetry audit techniques are required before these novel methodologies can be applied.