Half-normal saline versus normal saline for irrigation of open-irrigated radiofrequency catheters in atrial fibrillation ablation

J Cardiovasc Electrophysiol. 2021 Apr;32(4):973-981. doi: 10.1111/jce.14885. Epub 2021 Feb 26.

Abstract

Background: The creation of effective and permanent lesions is a crucial factor in determining the success rate of atrial fibrillation (AF) ablation. By increasing the efficacy of radiofrequency (RF) energy-mediated lesion formation, half-normal saline (HNS) as an irrigant for open-irrigated ablation catheters has the potential to reduce procedural times and improve acute and long-term outcomes.

Methods: This is a double-blind randomized clinical trial of 99 patients undergoing first-time RF catheter ablation for AF. Patients enrolled were randomly assigned in a 1:1 fashion to perform ablation using HNS or normal saline (NS) as an irrigant for the ablation catheter.

Results: The use of HNS is associated with shorter RF times (26 vs. 33 min; p = .02) with comparable procedure times (104 vs. 104 min). The rate of acute pulmonary vein reconnections (16% vs. 18%) was comparable, with a median of 1 vein reconnection in the HNS arm versus 2 in the NS arm. There was no difference in procedure-related complications, including the incidence of postprocedural hyponatremia when using HNS. Over the 1-year follow-up, there is no significant difference between the HNS and NS with respect to the recurrence of any atrial arrhythmia (off antiarrhythmic drugs [AAD]: 47% vs. 52%; hazard ratio [HR]: 1.17, 95% confidence interval [CI]: 0.66-2.06; off/on AAD: 66% vs. 66%, HR: 1.06, 95% CI: 0.53-2.12), with a potential benefit of using HNS when considering the paroxysmal AF cohort (on/off AAD 73% vs. 62%, HR: 0.72, 95% CI: 0.19-2.70).

Conclusions: In a mixed cohort of patients undergoing first-time AF ablation, irrigation of open-irrigated RF ablation catheters with HNS is associated with shorter RF times, with a comparably low rate of procedure-related complications. In the long term, there is no significant difference with respect to the recurrence of any atrial arrhythmia. Larger studies with a more homogeneous population are necessary to determine whether HNS improves clinical outcomes.

Keywords: atrial fibrillation ablation; half-normal saline; irrigated tip ablation catheter.

Publication types

  • Randomized Controlled Trial

MeSH terms

  • Atrial Fibrillation* / diagnosis
  • Atrial Fibrillation* / surgery
  • Catheter Ablation* / adverse effects
  • Catheters
  • Equipment Design
  • Humans
  • Pulmonary Veins* / surgery
  • Recurrence
  • Saline Solution
  • Treatment Outcome

Substances

  • Saline Solution