EFFECTS OF TEST-BOLUS AND LOW-DOSE SCAN ON CT PULMONARY ANGIOGRAPHY IMAGE QUALITY IN PATIENTS WITH DIFFERENT BODY MASS INDEXES

Radiat Prot Dosimetry. 2020 Dec 30;192(3):387-395. doi: 10.1093/rpd/ncaa217.

Abstract

This study aimed to investigate different methods of obtaining high-quality Computed Tomography pulmonary angiography (CTPA) images using low-dose scanning in patients with different body mass index (BMI) values. Sixty patients with suspected pulmonary embolism were grouped based on their BMI values (BMI < 25, designated N, and BMI ≥ 25, designated O) and were assigned to receive either test bolus (TB) or bolus tracking (BT) at conventional (C) or low (L) dose. The effective dose (ED) in the N-TB-L group was lower than in the group N-TB-C (0.56 ± 0.05 vs. 3.78 ± 1.16, p < 0.001), with similar image quality (4.90 ± 0.31 vs. 4.70 ± 0.47, p = 0.120). The ED in the O-TB-L group was lower than in the O-TB-C group (0.54 ± 0.03 vs. 5.14 ± 1.34, p < 0.001), but the group O-TB-C's image quality was higher (4.65 ± 0.59 vs. 3.95 ± 0.89, p = 0.006). Groups N-TB-L versus O-TB-L, groups N-TB-L versus N-BT-L and groups O-TB-C versus O-BT-C had similar EDs (all ps > 0.05), but the image quality was different (all ps < 0.05). In conclusion, the results showed that the image quality of low-dose CTPA scanning using TB was similar to that of the conventional-dose CTPA in patients with BMI < 25 but was lower in patients with BMI ≥ 25. TB was better than BT for all patients, regardless of BMI, when receiving the same ED.

MeSH terms

  • Angiography
  • Body Mass Index
  • Computed Tomography Angiography*
  • Contrast Media
  • Humans
  • Pulmonary Embolism* / diagnostic imaging
  • Radiation Dosage
  • Tomography, X-Ray Computed

Substances

  • Contrast Media