Abstract

Objectives: A prior publication introduced the Strome-Blitzer balloon's ability to obtain circumferential esophageal cytologic sampling. This GLP study was requisite for FDA approval to determine if equivalent cell capture and cellularity was observed with the balloon compared to surface sampling brushes and to determine the balloon's usability for naive otolaryngologists.

Methods: Three naïve users tested the Hobbs brush and Strome-Blitzer balloon on 4 Yorkshire swine. Four anatomical sites were sampled, beginning distally and ending proximally. In 2 animals, the balloon was used first distally and in the remaining 2, 4 new Hobbs brushes were used distally first. Moving proximally, the balloon and brushes were sequentially alternated. In follow-the-leader fashion, the balloon was introduced trans-orally followed by an endoscope to the desired site. The balloon was inflated exposing the abrasive strips to contact the esophageal mucosa. Moving the balloon 1 to 2 cm superiorly and inferiorly effected circumferential cell capture. The balloon was collapsed and removed, preserving the cellularity. The Hobbs brush was passed through the scope's channel. Four brushes, 1 per quadrant, obtained the samples at an anatomical site. The balloon was rated as pass/fail on the following: delivery, kinking, usability, and malfunction. A blinded veterinary pathologist evaluated the cytology.

Results: There was no device malfunction, mucosal trauma, or difficulty with device use. Balloon cytologic samples were comparable in cellularity and quality to the brush.

Conclusion: A single balloon sampling was comparable to 4 brushes in capturing diagnostically relevant cellular volumes and architecture. Naïve users easily performed the procedures after reading the guidelines.

Level of evidence: 3.

Keywords: Barrett’s esophagus; balloon cytology.