Patient Outcomes After Single-level Coflex Interspinous Implants Versus Single-level Laminectomy

Spine (Phila Pa 1976). 2021 Jul 1;46(13):893-900. doi: 10.1097/BRS.0000000000003924.

Abstract

Study design: Retrospective cohort analysis.

Objective: The aim of this study was to compare postoperative outcomes of Coflex interspinous device versus laminectomy.

Summary of background data: Coflex Interlaminar Stabilization device (CID) is indicated for one- or two-level lumbar stenosis with grade 1 stable spondylolisthesis in adult patients, as an alternative to laminectomy, or laminectomy and fusion. CID provides stability against progressive spondylolisthesis, retains motion, and prevents further disc space collapse.

Methods: Patients ≥18 years' old with lumbar stenosis and grade 1 stable spondylolisthesis who underwent either primary single-level decompression and implantation of CID, or single-level laminectomy alone were included with a minimum 90-day follow-up at a single academic institution. Clinical characteristics, perioperative outcomes, and postoperative complications were reviewed until the latest follow-up. χ2 and independent samples t tests were used for analysis.

Results: Eighty-three patients (2007-2019) were included: 37 cases of single-level laminectomy (48.6% female) were compared to 46 single-level CID (50% female). CID cohort was older (CID 69.0 ± 9.4 vs. laminectomy 64.2 ± 11.0, P = 0.042) and had higher American Society of Anesthesiologists (ASA) grade (CID 2.59 ± 0.73 vs. laminectomy 2.17 ± 0.48, P = 0.020). CID patients had higher estimated blood loss (EBL) (97.50 ± 77.76 vs. 52.84 ± 50.63 mL, P = 0.004), longer operative time (141.91 ± 47.88 vs. 106.81 ± 41.30 minutes, P = 0.001), and longer length of stay (2.0 ± 1.5 vs. 1.1 ± 1.0 days, P = 0.001). Total perioperative complications (21.7% vs. 5.4%, P = 0.035) and instrumentation-related complication was higher in CID (10.9% vs. 0% laminectomy group, P = 0.039). There were no other significant differences between the groups in demographics or outcomes.

Conclusion: Single-level CID devices had higher perioperative 90-day complications, longer operative time, length of stay, higher EBL compared to laminectomies alone. Similar overall revision and neurologic complication rates were noted compared to laminectomy at last follow-up.Level of Evidence: 3.

MeSH terms

  • Aged
  • Female
  • Humans
  • Laminectomy* / adverse effects
  • Laminectomy* / instrumentation
  • Male
  • Middle Aged
  • Postoperative Complications
  • Prostheses and Implants*
  • Retrospective Studies
  • Spondylolisthesis / surgery*
  • Treatment Outcome