Clinical validation of immunoassay HemosIL® AcuStar HIT-IgG (PF4-H) in the diagnosis of Heparin-induced thrombocytopenia

J Thromb Thrombolysis. 2021 Aug;52(2):601-609. doi: 10.1007/s11239-020-02349-4. Epub 2021 Jan 1.

Abstract

Heparin induced thrombocytopenia (HIT) is a life and limb-threatening complication of heparin exposure. The misdiagnosis of this disease can have major consequences on the patients. The objective of this study was to evaluate a diagnostic strategy that combines the 4Ts score with the result of HemosIL® AcuStar HIT-IgG (PF4-H) to confirm the diagnosis of HIT. Citrated plasmas from 1300 patients with suspicion of HIT were analyzed with a fully automated quantitative chemiluminescent immunoassay (HemosIL® AcuStar HIT-IgG (PF4/H)). If the IgG anti-PF4/H antibodies were positive (cut-off, 1 U/mL), HIT diagnosis was confirmed using functional tests. In total, 1300 samples of consecutive patients were enrolled, 94 (7.2%) of which gave positive results in HemosIL® AcuStar-IgG. HIT was diagnosed in 65 out of these patients, corresponding to a prevalence of 5%. Using ROC curve analysis, patients were divided into three groups according to their titer of antibodies. Higher values of the IgG (PF4-H) were associated with increased probability of HIT, and the diagnostic specificity was greatly increased using the combination of a 4Ts score > 3 and a positive titer ≥ 3.25 U/mL. Importantly, the diagnostic specificity is 100% when the titer is > 12.40 U/mL. We demonstrated that higher values of Anti PF4/H Antibodies were associated with a high probability of having HIT. A titer of HemosIL® IgG (PF4-H) > 12.40 U/mL has a specificity of 100% which should no require a functional test to confirm the diagnosis of HIT.

Keywords: 4 T score; Chemiluminescent; Heparin; Heparin induced thrombocytopenia; Immunoassay.

MeSH terms

  • Anticoagulants
  • Heparin / adverse effects
  • Humans
  • Immunoassay
  • Immunoglobulin G
  • Platelet Factor 4
  • ROC Curve
  • Thrombocytopenia* / chemically induced
  • Thrombocytopenia* / diagnosis

Substances

  • Anticoagulants
  • Immunoglobulin G
  • Platelet Factor 4
  • Heparin