Tezepelumab Pharmacokinetics, Safety, and Tolerability After Administration via Vial-and-syringe, Accessorized Prefilled Syringe, or Autoinjector: A Randomized Trial in Healthy Volunteers

Clin Ther. 2021 Jan;43(1):142-155.e5. doi: 10.1016/j.clinthera.2020.11.014. Epub 2020 Dec 27.

Abstract

Purpose: Tezepelumab is an anti-thymic stromal lymphopoietin monoclonal antibody therapeutic in development for patients with severe, uncontrolled asthma. In ongoing Phase III studies, tezepelumab is administered via subcutaneous (SC) injections using a vial-and-syringe (V-S). This study compared the pharmacokinetic (PK) parameters, safety, and tolerability of tezepelumab administered subcutaneously via V-S versus via an accessorized prefilled syringe (APFS) or autoinjector (AI).

Methods: This single-center, randomized, open-label, parallel-group study was conducted in healthy volunteers aged 18-65 years. Participants, stratified according to weight (50 to <70 kg, 70 to <80 kg, or 80-90 kg), were randomized evenly to 9 groups representing injections to the abdomen, thigh, or upper arm via V-S, APFS, or AI. Tezepelumab PK parameters over 113 days were evaluated after a single 210-mg SC dose. The primary end points were comparison of Cmax and AUC0-∞ between device groups. Further PK parameters, immunogenicity, safety (including injection site reactions [ISRs] and injection site pain [visual analog scale]) were also assessed.

Findings: A total of 315 adults were randomized to treatment. Geometric mean ratios for comparisons between device groups of Cmax, AUC0-∞, and AUC0-last were close to 1, with 90% CIs all within the range of 0.8-1.25, meeting bioequivalence criteria. PK variables were also similar between devices across injection sites and weight categories. Across devices, thigh injection resulted in slightly higher exposure than upper arm injection, and abdomen injection resulted in exposure similar to or slightly lower than thigh injection; however, these differences were not clinically meaningful. Treatment-emergent anti-tezepelumab antibodies were present in 3 (2.9%), 1 (1.0%), and 0 participants in the V-S, APFS, and AI groups, respectively. Treatment-related adverse events were reported in 15.0% of participants overall (V-S, 10.7%; APFS, 18.1%; AI, 16.0%), including ISRs in 1 (1.0%), 3 (2.9%), and 3 (2.8%) participants in the V-S, APFS, and AI groups. Median visual analog scale pain score (0-100 mm scale) was 2 mm immediately after injection and was 0 mm at 30 min for all groups.

Implications: Tezepelumab PK parameters after a single 210-mg SC dose were comparable when administered via V-S, APFS, or AI. In all groups, immunogenicity rate and injection site pain were low, and ISRs were uncommon. These findings support administration of tezepelumab via APFS or AI, in addition to V-S, providing patients and physicians with greater choice and the potential convenience of at-home use. ClinicalTrials.gov identifier: NCT03989544.

Keywords: asthma; bioequivalence; home use; injections; monoclonal antibody; thymic stromal lymphopoietin.

Publication types

  • Randomized Controlled Trial
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Anti-Asthmatic Agents / administration & dosage*
  • Anti-Asthmatic Agents / adverse effects
  • Anti-Asthmatic Agents / pharmacokinetics
  • Antibodies, Monoclonal, Humanized / administration & dosage*
  • Antibodies, Monoclonal, Humanized / adverse effects
  • Antibodies, Monoclonal, Humanized / pharmacokinetics
  • Drug Delivery Systems
  • Female
  • Healthy Volunteers
  • Humans
  • Injections, Subcutaneous
  • Male
  • Middle Aged
  • Pain Measurement
  • Syringes

Substances

  • Anti-Asthmatic Agents
  • Antibodies, Monoclonal, Humanized
  • tezepelumab

Associated data

  • ClinicalTrials.gov/NCT03989544