Retrieval of Long-Term Implanted Leadless Pacemakers: A Single-Center Experience

JACC Clin Electrophysiol. 2020 Dec;6(14):1744-1751. doi: 10.1016/j.jacep.2020.06.010. Epub 2020 Aug 26.

Abstract

Objectives: The authors report their single-center experience with the retrieval and replacement of long-term implanted leadless cardiac pacemakers (LPs) and the factors contributing to success.

Background: LPs are a clinically effective and safe alternative to standard transvenous pacemakers for single-chamber ventricular pacing. However, the feasibility of retrieving and replacing long-term implanted LPs is not well known.

Methods: A total of 34 patients with implanted Nanostim LPs (mean implantation duration 1,570 ± 479 days) subsequently underwent retrieval. On the basis of fluoroscopic imaging, the cohort was divided into 2 groups: those with remarkable swinging movement (SM) of the LP docking button (n = 25) and those without SM (n = 9).

Results: The overall LP retrieval success rate was 85% (n = 29). New leadless devices were implanted in 27 patients immediately after LP retrieval. No procedure-related adverse events occurred. The rate of successful LP retrieval was significantly higher in the SM group (25 of 25 [100%]) than in the no-SM group (4 of 9 [44%]) (p < 0.001), and fluoroscopy time during the retrieval procedure was shorter in the SM group (12.7 ± 8.6 min) than in the no-SM group (45.6 ± 19 min) (p < 0.001).

Conclusions: This study demonstrated the feasibility and safety of retrieval of long-term implanted LPs after a mean duration of 4 years. Retrieval was most successful in patients whose docking buttons exhibited significant SM.

Keywords: Nanostim; leadless cardiac pacemaker; pacemaker retrieval.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Equipment Design
  • Fluoroscopy
  • Humans
  • Pacemaker, Artificial* / adverse effects
  • Prostheses and Implants