Magnetic bead extraction with acid dissociation immunoassay for the determination of serum CD80-Fc fusion protein

Bioanalysis. 2021 Jan;13(2):89-100. doi: 10.4155/bio-2020-0259. Epub 2020 Dec 23.

Abstract

Background: To detect concentrations of subtherapeutic doses of the CD80-Fc fusion protein FPT155 in serum in Phase I studies, a highly sensitive assay was developed. Materials & methods: FPT155 was purified from human serum using magnetic beads coupled to cytotoxic T-lymphocyte-associated antigen-4. After washing away the serum components, FTP155 was released by acid dissociation and neutralization. The eluted drug was quantified in an ELISA using cytotoxic T-lymphocyte-associated antigen-4 as a capture reagent and biotinylated anti-human Fc for detection. The assay was validated with a calibration range of 5-40 ng/ml and a dilutional integrity of up to 100,000 ng/ml. Conclusion: A highly sensitive assay to determine serum concentrations of FPT155 using readily available reagents was developed. The results were in conformity with theoretical calculations.

Keywords: CD80-Fc fusion protein; Magnetic beads; acid dissociation; critical reagents; extraction; high-sensitivity assay.

MeSH terms

  • B7-1 Antigen / blood*
  • B7-1 Antigen / isolation & purification
  • Enzyme-Linked Immunosorbent Assay*
  • Humans
  • Hydrogen-Ion Concentration
  • Immunoglobulin Fc Fragments / blood*
  • Immunoglobulin Fc Fragments / isolation & purification
  • Magnetic Phenomena
  • Recombinant Fusion Proteins / blood*
  • Recombinant Fusion Proteins / isolation & purification

Substances

  • B7-1 Antigen
  • Immunoglobulin Fc Fragments
  • Recombinant Fusion Proteins