[Research of Methods to Reduce Alarm Fatigue of Monitoring System]

Zhongguo Yi Liao Qi Xie Za Zhi. 2020 Dec 8;44(6):481-486. doi: 10.3969/j.issn.1671-7104.2020.06.003.
[Article in Chinese]

Abstract

Objective: In order to solve alarm fatigue, the algorithm optimization strategies were researched to reduce false and worthless alarms.

Methods: A four-lead arrhythmia analysis algorithm, a multiparameter fusion analysis algorithm, an intelligent threshold reminder, a refractory period delay technique were proposed and tested with collected 28 679 alarms in multi-center study.

Results: The sampling survey indicate that the 80.8% of arrhythmia false alarms were reduced by the four-lead analysis, the 55.9% of arrhythmia and pulse false alarms were reduced by the multi-parameter fusion analysis, the 28.0% and 29.8% of clinical worthless alarms were reduced by the intelligent threshold and refractory period delay techniques respectively. Finally, the total quantity of alarms decreased to 12 724.

Conclusions: To increase the dimensionality of parametric analysis and control the alarm limits and delay time are conducive to reduce alarm fatigue in intensive care units.

Keywords: alarm fatigue; critical care; intelligent threshold; multi-parameter fusion; nursing management.