Retinal safety evaluation of photoacoustic microscopy

Exp Eye Res. 2021 Jan:202:108368. doi: 10.1016/j.exer.2020.108368. Epub 2020 Nov 24.

Abstract

Photoacoustic microscopy (PAM) has significant potential as a promising diagnostic method for eye diseases and can provide anatomic and functional information of the retinal and choroidal vasculature. However, there are no FDA-approved PAM systems for ophthalmic imaging. In this study, a comprehensive safety evaluation was performed to evaluate the safety of PAM retinal imaging and whether PAM causes damage to retinal structure or function in rabbit eyes. 12 Dutch-Belted pigmented rabbits received photoacoustic imaging to 57% of the retinal surface area with a laser energy of 5% of the ANSI safety limit for five consecutive days and followed before imaging and 3 days, 1, 2, 3, and 4 weeks post imaging. Retinal morphologic analyses using slit lamp examination, fundus photography, red free, FA, FAF, ICGA, and OCT showed no retinal hemorrhage, edema, detachment, vascular abnormalities, or pigmentary abnormalities in the retina or choroid after PAM imaging. Full-field ERG analysis showed no significant difference in scotopic or photopic a- and b-wave amplitudes or implicit times between the control and experimental eyes over time (n = 6, P values > 0.05). Retinal ultrastructural evaluation using TEM showed normal structure of organelles and nuclei, and no significant loss of cells after PAM. TUNEL assay showed no evidence of cells apoptosis in retina. Retinal histopathology indicated that the architecture and thickness of the retinal layers was well preserved in all experimental eyes. A positive control at 500% of the ANSI limit demonstrated significant damage. The comprehensive retinal safety evaluation demonstrated no damage to retinal structure or function for 4 weeks after PAM imaging in rabbits.

Keywords: ERG; Photoacoustic microscopy; Retinal imaging; Safety evaluation.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Microscopy, Acoustic / methods*
  • Models, Animal
  • Photoacoustic Techniques / methods*
  • Rabbits
  • Reproducibility of Results
  • Retina / diagnostic imaging*
  • Retinal Vessels / diagnostic imaging