Mechanisms for the Clinical Utility of Low-Frequency Stimulation in Neuromodulation of the Dorsal Root Ganglion

Neuromodulation. 2021 Jun;24(4):738-745. doi: 10.1111/ner.13323. Epub 2020 Nov 25.

Abstract

Background: Dorsal root ganglion stimulation (DRG-S) involves the electrical modulation of the somata of afferent neural fibers to treat chronic pain. DRG-S has demonstrated clinical efficacy at frequencies lower than typically used with spinal cord stimulation (SCS). In a clinical study, we found that the frequency of DRG-S can be tapered to a frequency as low as 4 Hz with no loss of efficacy. This review discusses possible mechanisms of action underlying effective pain relief with very low-frequency DRG-S.

Materials and methods: We performed a literature review to explore the role of frequency in neural transmission and the corresponding relevance of frequency settings with neuromodulation.

Findings: Sensory neural transmission is a frequency-modulated system, with signal frequency determining which mechanisms are activated in the dorsal horn. In the dorsal horn, low-frequency signaling (<20 Hz) activates inhibitory processes while higher frequencies (>25 Hz) are excitatory. Physiologically, low-threshold mechanoreceptors (LTMRs) fibers transmit or modulate innocuous mechanical touch at frequencies as low as 0.5-5 Hz, while nociceptive fibers transmit pain at high frequencies. We postulate that very low-frequency DRG-S, at least partially, harnesses LTMRs and the native endogenous opioid system. Utilizing lower stimulation frequency decreases the total energy delivery used for DRG-S, extends battery life, and facilitates the development of devices with smaller generators.

Keywords: Dorsal root ganglion; frequency; low back pain; neuromodulation; pathway; stimulation; transmission.

Publication types

  • Review

MeSH terms

  • Chronic Pain* / therapy
  • Ganglia, Spinal
  • Humans
  • Nerve Fibers
  • Pain Management
  • Spinal Cord Stimulation*