Low-Profile Induced-Voltage Distance Ranger for Smart Contact Lenses

IEEE Trans Biomed Eng. 2021 Jul;68(7):2203-2210. doi: 10.1109/TBME.2020.3040161. Epub 2021 Jun 17.

Abstract

Objective: In this paper, we present a novel, low-profile, scleral-coil based, distance ranging system which is suitable for smart, accommodating contact lenses.

Methods: We measure the induced emf between a set of four thin semi-circular coils patterned on flexible Kapton substrates that conform to the eyes' sclera. This induced emf is a function of eye gaze angles. The system then determines the distance from the eyes to the desired object via the triangulation of these eye gaze angles Results: Experiments on eyeball simulated tissue gels indicate an accurate prediction of object distance in the 0.1-15 D range with a 0.15 D RMS error and object direction in the -15 to 15-degree arc with 0.4-degree RMS error, respectively. The energy required was determined to be as low as 20 μJ per range reading.

Conclusion: Experimental data shows that our proposed new method of eye-tracking and distance ranging system can accurately predict eye-gaze angles and object-distance, whilst using only 20 μJ per range reading.

Significance: The high-accuracy, low-profile and reduced energy requirements of the proposed eye-tracking technique, make it suitable for applications in the vast field of adaptive optics such as smart contact lenses and other low-power vision corrective applications.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, U.S. Gov't, Non-P.H.S.

MeSH terms

  • Contact Lenses*
  • Fixation, Ocular
  • Sclera