Long-Term Stability and Safety of the Soundbridge Coupled to the Round Window

Laryngoscope. 2021 May;131(5):E1434-E1442. doi: 10.1002/lary.29269. Epub 2020 Nov 19.

Abstract

Objective: The objective of the study was to demonstrate the long-term outcomes of patients implanted with the active middle ear implant (AMEI) Vibrant Soundbridge (VSB) through coupling the floating mass transducer (FMT) to the round window (RW).

Methods: This retrospective study evaluated the short- and long-term clinical performance (audiological outcomes) and safety (revisions/explantations) of the VSB coupled to the RW between 2013 and 2019 at the St. Pölten University Hospital, Austria. For the outcome analysis, the sample was divided into a short-term examination group followed up for less than 12 months (<12 months) and a long-term examination group followed up for more than 12 months (>12 months). Cumulative survival outcomes were separately analyzed for subjects with and without cholesteatoma.

Results: 46 patients with an average long-term follow-up period of 31.43 months (13-75 months) were investigated. Complications requiring revision surgery were reported in total in seven patients with cholesteatoma (15.2%) and none in subjects without cholesteatoma (0%). Residual hearing was not affected by VSB surgery. Word understanding on the Freiburger monosyllabic speech test improved significantly at 65 dB (P < .0001) and 80 dB (P < .0001), and these outcomes were stable for long-term follow up.

Conclusion: The VSB coupled to the RW is a safe implantation method for patients with conductive or mixed hearing loss. Hearing improvement was stable for the long-term follow-up up to 74 months. The revision rates are directly related to the underlying pathology of cholesteatoma (with radical cavity); thus, this special cohort requires additional counseling on potential complications.

Level of evidence: 4 (Case-series) Laryngoscope, 131:E1434-E1442, 2021.

Keywords: AMEI; Round window; VSB; active middle ear implant; follow-up; implantation; long-term; performance; vibrant soundbridge.

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Austria / epidemiology
  • Cholesteatoma, Middle Ear / complications
  • Cholesteatoma, Middle Ear / surgery*
  • Device Removal / statistics & numerical data
  • Female
  • Follow-Up Studies
  • Hearing Loss, Conductive / diagnosis
  • Hearing Loss, Conductive / etiology
  • Hearing Loss, Conductive / surgery*
  • Hearing Loss, Mixed Conductive-Sensorineural / diagnosis
  • Hearing Loss, Mixed Conductive-Sensorineural / etiology
  • Hearing Loss, Mixed Conductive-Sensorineural / surgery*
  • Humans
  • Male
  • Middle Aged
  • Ossicular Prosthesis / adverse effects
  • Ossicular Replacement / adverse effects*
  • Ossicular Replacement / instrumentation
  • Ossicular Replacement / methods
  • Postoperative Complications / epidemiology*
  • Postoperative Complications / etiology
  • Retrospective Studies
  • Round Window, Ear / surgery*
  • Treatment Outcome
  • Young Adult