Comparison of image quality and radiation dose using different pre-ASiR-V and post-ASiR-V levels in coronary computed tomography angiography

J Xray Sci Technol. 2021;29(1):125-134. doi: 10.3233/XST-200754.

Abstract

Objective: To determine the optimal pre-adaptive and post-adaptive level statistical iterative reconstruction V (ASiR-V) for improving image quality and reducing radiation dose in coronary computed tomography angiography (CCTA).

Methods: The study was divided into two parts. In part I, 150 patients for CCTA were prospectively enrolled and randomly divided into 5 groups (A, B, C, D, and E) with progressive scanning from 40% to 80% pre-ASiR-V with 10% intervals and reconstructing with 70% post-ASiR-V. The signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were calculated. Subjective image quality was assessed using a 5-point scale. The CT dose index volume (CTDIvol) and dose-length product (DLP) of each patient were recorded and the effective radiation dose (ED) was calculated after statistical analysis by optimizing for the best pre-ASiR-V value with the lowest radiation dose while maintaining overall image quality. In part II, the images were reconstructed with the recommended optimal pre-ASiR-V values in part I (D group) and 40%-90% of post-ASiR-V. The reconstruction group (D group) was divided into 6 subgroups (interval 10%, D0:40% post-ASiR-V, D1:50% post - ASiR-V, D2:60% post-ASiR-V, D3:70% post-ASiR-V, D4:80% post-ASiR-V, and D5:90% post-ASiR-V).The SNR and CNR of D0-D5 subgroups were calculated and analyzed using one-way analysis of variance, and the consistency of the subjective scores used the k test.

Results: There was no significant difference in the SNRs, CNRs, and image quality scores among A, B, C, and D groups (P > 0.05). The SNR, CNR, and image quality scores of the E group were lower than those of the A, B, C, and D groups (P < 0.05). The mean EDs in the B, C, and D groups were reduced by 7.01%, 13.37%, and 18.87%, respectively, when compared with that of the A group. The SNR and CNR of the D4-D5 subgroups were higher than the D0-D3 subgroups, and the image quality scores of the D4 subgroups were higher than the other subgroups (P < 0.05).

Conclusion: The wide-detector combined with 70% pre-ASiR-V and 80% post-ASiR-V significantly reduces the radiation dose of CCTA while maintaining overall image quality as compared with the manufacture's recommendation of 40% pre-ASiR-V.

Keywords: CCTA; Post-ASiR-V; Pre-ASiR-V; image quality; radiation dose.

Publication types

  • Randomized Controlled Trial
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Algorithms
  • Computed Tomography Angiography*
  • Humans
  • Radiation Dosage
  • Radiographic Image Interpretation, Computer-Assisted*
  • Signal-To-Noise Ratio
  • Tomography, X-Ray Computed