Enhanced spatial resolution of Eljen-204 plastic scintillators for use in rep-rated proton diagnostics

Rev Sci Instrum. 2020 Oct 1;91(10):103301. doi: 10.1063/5.0014949.

Abstract

A pixelated scintillator has been designed, fabricated, and tested using a laser-accelerated proton source for use in proton diagnostics at rep-rated laser facilities. The work presented here demonstrates the enhanced spatial resolution of thin, organic scintillators through a novel pixelation technique. Experimental measurements using laser-generated protons incident onto 130 μm-thick scintillators indicate a >20% reduction in the scintillator point spread function (PSF) for the detectors tested. The best performing pixelated detector reduced the ∼200 μm PSF of the stock material to ∼150 μm. The fabrication technique may be tailored to reduce the pixel size and achieve higher spatial resolutions.