Usefulness of handheld electroretinogram system for diagnosing blue-cone monochromatism in children

Jpn J Ophthalmol. 2021 Jan;65(1):23-29. doi: 10.1007/s10384-020-00782-9. Epub 2020 Nov 2.

Abstract

Purpose: To report the diagnosis of three childhood patients with blue-cone monochromatism (BCM) using S-cone electroretinograms (ERG) recorded with RETeval® Complete.

Study design: Prospective clinical study.

Methods: We examined three boys initially suspected of having rod monochromatism. S-cone ERG was performed with red background and blue flashed light stimulation using two different intensities: 0.25 cd × s/m2 and 1 cd × s/m2.

Results: Case 1 was a 12-year-old boy with a visual acuity of 0.1 OU. Case 2 was an 8-year-old boy with a visual acuity of 0.3 OD and 0.2 OS. Both cases showed a myopic fundus and nystagmus without any other ocular abnormalities. Case 3 was a 6-year-old boy with a visual acuity of 0.3 OD and 0.4 OS. He also showed myopic fundus changes, but nystagmus was not observed. Rod and maximal responses recorded with RETeval® were likely to be within normal range; however, cone responses were absent in all cases. S-cone ERGs showed positive responses at 40 ms with 0.25 cd × s/m2 intensity in Case 2, and at approximately 30-40 ms with 1.0 cd × s/m2 intensity in all three cases. These ERG findings led to a diagnosis of BCM.

Conclusions: S-cone ERG of RETeval® was helpful in diagnosing with minimal invasion BCM in childhood patients.

Keywords: Blue-cone monochromatism; Reteval®; Retinal disease; S-cone electroretinograms.

Publication types

  • Case Reports

MeSH terms

  • Child
  • Color Vision Defects* / diagnosis
  • Electroretinography
  • Humans
  • Male
  • Prospective Studies
  • Retinal Cone Photoreceptor Cells