Early bioprosthesis failure in mitral valve replacement

J Card Surg. 2021 Jan;36(1):307-308. doi: 10.1111/jocs.15165. Epub 2020 Oct 30.

Abstract

The causes of early bioprosthesis failure include infective endocarditis, pannus formation, and structural valve deterioration. We reported a patient who suffered from early mitral bioprosthesis failure due to leaflets restricted by the subvalvular apparatus and early pannus formation. In patients with symptoms relapse and mitral regurgitation recurrence early after mitral valve replacement, early pannus formation needs to be anticipated, and surgical intervention should be performed if symptoms persist after medical treatment.

Keywords: bioprosthesis failure; valve repair/replacement.

MeSH terms

  • Bioprosthesis*
  • Heart Valve Prosthesis Implantation*
  • Heart Valve Prosthesis*
  • Humans
  • Mitral Valve / diagnostic imaging
  • Mitral Valve / surgery
  • Mitral Valve Insufficiency* / etiology
  • Mitral Valve Insufficiency* / surgery
  • Prosthesis Failure
  • Reoperation