Rapid Detection of Femtogram Amounts of Protein by Gel-Free Immunoblot

Bull Exp Biol Med. 2020 Oct;169(6):840-843. doi: 10.1007/s10517-020-04988-2. Epub 2020 Oct 24.

Abstract

The article presents a new method of immunoblotting for simple, rapid, and highly sensitive detection of proteins. Electrophoretic separation of sample is carried out under non-denaturing conditions in a thin conductive layer between cellulose membranes without polyacrylamide gel. The membrane surface is preliminarily modified with azidophenyl groups to photochemically immobilize proteins in situ. For visualization of protein bands, the membranes are treated with magnetic beads coated with specific antibodies, unbound particles are then removed with a magnet. The detection limit in the model system with biotinylated BSA and magnetic beads coated with streptavidin reaches 10 fg or about 105 molecules, while the total blotting time does not exceed 5 min. The method was applied for detection of IgA in a sample of human exhaled air. The method can be used for the analysis of various complex biological samples containing low amounts of the analyte.

Keywords: electrophoresis; immunoblotting; immunoglobulin A; magnetic beads.

MeSH terms

  • Air / analysis
  • Azides / chemistry
  • Biotin / chemistry
  • Biotinylation
  • Cellulose / chemistry
  • Electrophoresis / instrumentation
  • Electrophoresis / methods*
  • Exhalation / physiology
  • Humans
  • Immobilized Proteins / analysis*
  • Immunoblotting / instrumentation
  • Immunoblotting / methods*
  • Immunoglobulin A / analysis*
  • Immunomagnetic Separation / methods*
  • Limit of Detection
  • Membranes, Artificial
  • Photochemical Processes
  • Serum Albumin, Bovine / chemistry
  • Streptavidin / chemistry

Substances

  • Azides
  • Immobilized Proteins
  • Immunoglobulin A
  • Membranes, Artificial
  • Serum Albumin, Bovine
  • Biotin
  • Cellulose
  • Streptavidin