A gadolinium-based magnetic ionic liquid for dispersive liquid-liquid microextraction

Anal Bioanal Chem. 2021 Jan;413(1):205-214. doi: 10.1007/s00216-020-02992-z. Epub 2020 Oct 23.

Abstract

A hydrophobic gadolinium-based magnetic ionic liquid (MIL) was investigated for the first time as an extraction solvent in dispersive liquid-liquid microextraction (DLLME). The tested MIL was composed of trihexyl(tetradecyl)phosphonium cations and paramagnetic gadolinium chloride anions. The prepared MIL showed low water miscibility, reasonable viscosity, markedly high magnetic susceptibility, adequate chemical stability, low UV background, and compatibility with reversed-phase HPLC solvents. These features resulted in a more efficient extraction than the corresponding iron or manganese analogues. Accordingly, the overall method sensitivity and reproducibility were improved, and the analysis time was reduced. The applicability of the proposed MIL was examined through the microextraction of four sartan antihypertensive drugs from aqueous samples followed by reversed-phase HPLC with UV detection at 240 nm. The DLLME procedures were optimized for disperser solvent type, MIL mass, disperser solvent volume, as well as acid, base, and salt addition. The limits of quantitation (LOQs) obtained with the analysis of 1.2-mL samples after DLLME and HPLC were 80, 30, 40, and 160 ng/mL for azilsartan medoxomil, irbesartan, telmisartan, and valsartan, respectively. Correlation coefficients were greater than 0.9988 and RSD values were in the range of 2.48-4.07%. Under the optimized microextraction conditions and using a 5-mL sample volume, enrichment factors were raised from about 40 for all sartans using a 1.2-mL sample to 175, 176, 169, and 103 for azilsartan medoxomil, irbesartan, valsartan, and telmisartan, respectively. The relative extraction recoveries for the studied sartans in river water varied from 82.5 to 101.48% at a spiked concentration of 0.5 μg/mL for telmisartan and irbesartan and 1 μg/mL for azilsartan medoxomil and valsartan. Graphical abstract.

Keywords: Dispersive liquid–liquid microextraction; High-performance liquid chromatography; Magnetic ionic liquids; Sartans.

MeSH terms

  • Chromatography, High Pressure Liquid
  • Gadolinium / chemistry*
  • Humans
  • Hydrophobic and Hydrophilic Interactions
  • Ionic Liquids / chemistry*
  • Ionic Liquids / toxicity
  • Limit of Detection
  • Liquid Phase Microextraction / methods*
  • Magnetics*
  • Reproducibility of Results
  • Solvents / chemistry
  • Spectrophotometry, Ultraviolet
  • Water / chemistry

Substances

  • Ionic Liquids
  • Solvents
  • Water
  • Gadolinium