Simpler and effective radiological evaluations for modiolar proximity of a slim modiolar cochlear implant electrode

Sci Rep. 2020 Oct 19;10(1):17714. doi: 10.1038/s41598-020-74738-x.

Abstract

A new slim modiolar electrode (CI532/632) has been reported to ensure better modiolar proximity than conventional electrodes. Better modiolar proximity has been proposed to yield better electrode discrimination capability and potentially better speech outcomes, necessitating its efficient measurement. Currently, intracochlear positional index (ICPI), the most reliable indicator for evaluating modiolar proximity, has been measured exclusively through 'metal artifact-less' cone beam CT. However, popular use of this index is precluded due to lack of cone beam CT in many institutions. Thus, eyes are now on elucidation of easy-to-measure indicators of modiolar proximity derived from conventional CT, which is accessible in all centers. We observed that enhanced tomographic resolution significantly reduces partial volume artifacts, providing better visualization of modiolus-electrode distance. Aided by ultra-high kernel specification with high-resolution index, we developed a novel and easy-to-measure, conventional CT-specific indicator, "modified ICPI", for evaluation of modiolar proximity. Further, we showed that it closely correlates with the previously proposed parameter of modiolar proximity, the spiral diameter, measured from post-insertion radiograph, reiterating the value of X-ray-based spiral diameter. Through this study, we have taken a step toward the stage of immediate visual feedback regarding modiolar proximity and changes in insertion technique intraoperatively, ensuring optimal modiolar proximity.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Artifacts
  • Cochlear Implantation / methods*
  • Cochlear Implants*
  • Electrodes*
  • Equipment Design*
  • Humans
  • Tomography, X-Ray Computed / methods*