Comparison of the reliability of Gram-negative and Gram-positive flags of the Sysmex UF-5000 with manual Gram stain and urine culture results

Clin Chem Lab Med. 2020 Oct 12;59(3):619-624. doi: 10.1515/cclm-2020-1263. Print 2021 Feb 23.

Abstract

Objectives: Recently, the fully automated flow cytometry-based UF-5000 (Sysmex Corboration, Kobe, Japan) urine sediment analyzer was developed providing bacteria (BACT) info flags for more accurate bacterial discrimination of urinary tract infections (UTIs). This study aimed to compare the reliability of the UF-5000 BACT-info flags with manual Gram stain and urine culture as the gold standard method.

Methods: A total of 344 urine samples were analyzed on the UF-5000 and compared with manual microscopic Gram stain and urine cultures. Agreement was assessed by Cohen's kappa (κ) analysis. The Youden index was used to determine the optimal BACT and white blood cell (WBC) cut-off points for discriminating positive and negative urine cultures.

Results: Overall 98/344 (28.5%) samples were urine culture positive at a cut-off of ≥105 CFU/mL. "Gram-negative?" UF-5000 BACT-Info flags showed a better concordance of 25/40 (62.5%) with urine culture compared to Gram stain with 30/50 (60%). The results for UF-5000 discrimination of Gram-positive and Gram-negative microorganisms demonstrated a substantial (κ = 0.78) and fair (κ = 0.40) agreement with urine culture. Optimal cut-off points detecting positive urine cultures were 135 BACT/µL (sensitivity [SE]: 92.1%, specificity [SP]: 85.4%, positive predictive value [PPV]: 71%, negative predictive value [NPV]: 96%) and 23 WBC/µL (SE: 73.5%, SP: 84.1%, PPV: 65%, NPV: 89%).

Conclusions: The UF-5000 analyzer (Sysmex) is a reliable diagnostic tool for UTI screening. The displayed BACT-Info flags allow a quick diagnostic orientation for the clinician. However, the authors suggest verifying the automated Gram categories with urine culture.

Keywords: automation; bacteriuria; urinalysis; urinary tract infections.

MeSH terms

  • Bacteria
  • Flow Cytometry
  • Humans
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Urinalysis
  • Urinary Tract Infections* / diagnosis
  • Urine