Comparison of the rotational stability between plate-haptic toric and C-loop haptic toric IOLs in myopic eyes

J Cataract Refract Surg. 2020 Oct;46(10):1353-1359. doi: 10.1097/j.jcrs.0000000000000259.

Abstract

Purpose: To compare the rotational stability of a plate-haptic toric intraocular lens (IOL) vs a C-loop haptic toric IOL in myopic cataract eyes.

Setting: Eye and Ear, Nose, and Throat Hospital of Fudan University, China.

Design: Prospective, randomized, controlled study.

Methods: Cataract eyes with axial length (AL) more than 24.5 mm were randomly assigned to receive implantation of a C-loop haptic toric IOL (AcrySof Toric IOL) (Group A) or a plate-haptic toric IOL (AT TORBI 709M IOL) (Group B). IOL rotation, residual astigmatism, visual acuity, and higher-order aberrations (HOAs) evaluated with OPD-Scan III aberrometer were compared at 3 months postoperatively.

Results: In total, 62 eyes of 62 patients were eligible for analysis: 31 in Group A and 31 in Group B. The mean rotation of toric IOLs was greater in Group A than that in Group B (8.00 ± 3.60 degrees vs 4.42 ± 3.24 degrees, respectively, P < .001), especially when IOLs were vertically placed. IOL rotation was positively correlated with AL in Group A, whereas no such correlations were found in Group B. Residual astigmatism in Group A was greater than that in Group B (-0.76 ± 0.30 diopter [D] vs -0.51 ± 0.29 D, respectively, P = .001). Fewer eyes achieved residual astigmatism of 0.50 D or less in Group A than in Group B (38.71% vs 64.52%). Group A had worse postoperative uncorrected visual acuity and higher total HOAs and coma for a 6.0 mm pupil than Group B, whereas postoperative corrected visual acuity was not different between the 2 groups.

Conclusions: The plate-haptic toric IOL might be a better choice for myopic cataract eyes with corneal astigmatism because of reduced postoperative rotation.

Publication types

  • Randomized Controlled Trial

MeSH terms

  • Astigmatism* / surgery
  • Cataract*
  • China
  • Humans
  • Lens Implantation, Intraocular
  • Lenses, Intraocular*
  • Phacoemulsification*
  • Prospective Studies
  • Refraction, Ocular