A Two-Electrode, Double-Pulsed Sensor Readout Circuit for Cardiac Troponin I Measurement

IEEE Trans Biomed Circuits Syst. 2020 Dec;14(6):1362-1370. doi: 10.1109/TBCAS.2020.3029912. Epub 2020 Dec 31.

Abstract

This paper presents a pulse-stimulus sensor readout circuit for use in cardiovascular disease examinations. The sensor is based on a gold nanoparticle plate with an antibody post-modification. The proposed system utilizes gated pulses to detect the biomarker Cardiac Troponin I in an ionic solution. The characteristic of the electrostatic double-layer capacitor generated by the analyte is related to the concentration of Cardiac Troponin I in the solvent. After sensing by the transistor, a current-to-frequency converter (I-to-F) and delay-line-based time-to-digital converter (TDC) convert the information into a series of digital codes for further analysis. The design is fabricated in a 0.18-μm standard CMOS process. The chip occupies an area of 0.92 mm2 and consumes 125 μW. In the measurements, the proposed circuit achieved a 1.77 Hz/pg-mL sensitivity and 72.43 dB dynamic range.

MeSH terms

  • Biosensing Techniques* / instrumentation
  • Biosensing Techniques* / methods
  • Electrodes
  • Equipment Design
  • Gold / chemistry
  • Humans
  • Metal Nanoparticles / chemistry
  • Semiconductors
  • Troponin I / analysis*
  • Troponin I / blood

Substances

  • Troponin I
  • Gold