Selective Wireless Stimulation of Rat Sciatic Nerve

Annu Int Conf IEEE Eng Med Biol Soc. 2020 Jul:2020:3407-3410. doi: 10.1109/EMBC44109.2020.9175335.

Abstract

Chronic stability of functional performance is a significant challenge to the success of implantable devices for neural stimulation and recording. Integrating wireless technology with typical microelectrode array designs is one approach that may reduce instances of mechanical failure and improve the long-term performance of neural devices. We have investigated the long-term stability of Wireless Floating Microelectrode Arrays (WMFAs) implanted in rat sciatic nerve, and their ability to selectively recruit muscles in the hind limb via neural stimulation. Thresholds as low as 4.1 μA were able to generate visible motion of the rear paw. Each implanted device (n=6) was able to selectively recruit plantar flexion and dorsiflexion of the rear paw, and selective stimulation of both movements was achieved throughout the study period. The evoked limb motion was electrode specific and was dependent on location within the fascicular structure of the nerve. Motor thresholds and movement patterns remained stable for more than 8 weeks after device implantation. No major changes in limb function were observed between the implanted and contralateral limb, or between implanted animals and control group animals. The results of this study show that WFMAs with intrafascicular electrodes implanted in a healthy peripheral nerve can provide stable and selective motor recruitment, without altering overall limb function.

MeSH terms

  • Animals
  • Hindlimb
  • Microelectrodes
  • Movement
  • Rats
  • Sciatic Nerve*
  • Wireless Technology*