A Novel Method for Measuring Serum Unbound Bilirubin Levels Using Glucose Oxidase-Peroxidase and Bilirubin-Inducible Fluorescent Protein (UnaG): No Influence of Direct Bilirubin

Int J Mol Sci. 2020 Sep 16;21(18):6778. doi: 10.3390/ijms21186778.

Abstract

The glucose oxidase-peroxidase (GOD-POD) method used to measure serum unbound bilirubin (UB) suffers from direct bilirubin (DB) interference. Using a bilirubin-inducible fluorescent protein from eel muscle (UnaG), a novel GOD-POD-UnaG method for measuring UB was developed. Newborn sera with an indirect bilirubin/albumin (iDB/A) molar ratio of <0.5 were classified into four groups of DB/total serum bilirubin (TB) ratios (<5%, 5-10%, 10-20%, and ≥20%), and the correlation between the UB levels and iDB/A ratio was examined. Linear regression analysis was performed to compare UB values from both methods with the iDB/A ratio from 38 sera samples with DB/TB ratio <5% and 11 samples with DB/TB ratio ≥5%. The correlation coefficient (r) between UB values and the iDB/A ratio for the GOD-POD method was 0.8096 (DB/TB ratio <5%, n = 239), 0.7265 (5-10%, n = 29), 0.7165 (10-20%, n = 17), and 0.4816 (≥20%, n = 16). UB values using the GOD-POD-UnaG method highly correlated with the iDB/A ratio in both <5% and ≥5% DB/TB ratio sera (r = 0.887 and 0.806, respectively), whereas a low correlation (r = 0.428) occurred for ≥5% DB/TB ratio sera using the GOD-POD method. Our GOD-POD-UnaG method can measure UB levels regardless of the presence of DB.

Keywords: albumin; analyzer; bilirubin encephalopathy; glucose oxidase-peroxidase method; indirect bilirubin; neonatal intensive care unit.

Publication types

  • Comparative Study
  • Validation Study

MeSH terms

  • Artifacts
  • Bilirubin / blood*
  • Equipment Design
  • Fetal Blood / chemistry*
  • Gestational Age
  • Glucose Oxidase
  • Humans
  • Hyperbilirubinemia, Neonatal / blood*
  • Liver Function Tests / instrumentation
  • Liver Function Tests / methods*
  • Oxidation-Reduction
  • Peroxidase
  • Reproducibility of Results
  • Retrospective Studies
  • Serum / chemistry
  • Spectrometry, Fluorescence / instrumentation
  • Spectrometry, Fluorescence / methods
  • Spectrophotometry / instrumentation
  • Spectrophotometry / methods

Substances

  • Glucose Oxidase
  • Peroxidase
  • Bilirubin